·双层玻璃反应釜的常见故障

1, vacuum hose aging: please replacethe vacuum hose. When using the the Great Wall instrument glass reactionkettle, the power switch is turned on, and the indicator light is not bright:the external power supply is not connected or the contact is bad;

2, the power indicator light is bright,but not rotation: rotating shaft rust, stop using, contact with suppliers.Motor, electric box failure, not connected to the seven core plug, re connectthe seven core plug;

3, a vacuum, but pumping: abrasion ofthe sealing ring, connected to a vacuum switch leakage please replace thesealing ring and a switch. The vacuum from time to time: steel shaft there isdirt connector is loose, please remove dirt check the vacuum gauge, a vacuumpump;

5, the insurance tube short: will thepower switch in the OFF position, and then change the insurance tube;